Do you offer CMM calibration near me?
Service covers Milwaukee and the Wisconsin metros listed on this site, including Waukesha, Racine, Sheboygan, Fond du Lac, Oshkosh and Green Bay, generally as on-site visits for fixed machines. Outside that radius,
to confirm scheduling and travel.What's the difference between calibration and verification?
Calibration establishes the relationship between a machine's readings and a reference value, with stated uncertainty. Most work sold as CMM calibration is actually performance verification against a specification such as ISO 10360-2.
What happens if the machine fails as-found?
As-found data is recorded before adjustment. Out-of-tolerance results are reported with enough detail to support an impact assessment back to the last passing calibration, before any geometric adjustment proceeds.
What does ISO 10360-2 actually test?
Length measurement error, E0 and E150, is measured at five calibrated lengths, three repeats each, across seven positions including the four space diagonals: 105 measurements compared against E0,MPE.
How long is the machine down for a verification visit?
Time on site depends on machine size and axis travel; large-envelope bridge and gantry machines common to machinery builders take longer than a small production CMM.
for a time-on-site estimate.What temperature does the room need to be for testing?
ISO 1 fixes 20°C as the dimensional reference temperature. For acceptance testing the machine manufacturer's specified environment applies; for reverification, the environment is recorded on the certificate regardless of who specifies it.
Is a portable arm covered on the same certificate as a bridge CMM?
Bridge and gantry machines are tested under ISO 10360-2 or ASME B89.4.1. Articulated arms are tested separately under ISO 10360-12:2016, where applicable to the arm and probe configuration, with ASME B89.4.22 as the US analogue; orders written around European-built machines may instead cite VDI/VDE 2617.
What is a CMC and why does it matter to my tolerances?
CMC, calibration and measurement capability, is the smallest uncertainty a laboratory can claim within its accredited scope. The reported uncertainty is checked against that scope, and against the tolerance being inspected, before a result is issued.
Will this satisfy my IATF or AS9100 auditor?
IATF 16949 and AS9100 both expect evidence that an external laboratory holds ISO/IEC 17025 accreditation or is otherwise acceptable to the customer. Certificates issued under a laboratory's accredited scope are built to carry that evidence.
What is an error map, and can it be updated on my controller?
The error map is the controller's volumetric compensation table. It can be updated on most controllers during geometric adjustment; some OEM controllers lock this map and require vendor-specific access.
Further Detail
Standards, traceability and equipment control are set out on the quality system page. Competence and authorization are covered on the our team page. Commercial terms are in the terms and conditions.